Measuring Instruments

Product Selection

The WGA-910A, WGA-680A, and WP-120A support CC-Link.
The WP-120A supports EtherNet/IPTM.

  • Please see here for more information about WGA-910A.
  • Please see here for more information about WGA-680A.
  • Please see here for more information about WP-120A.

Primarily, there are four types of measuring instruments which are

STRAIN AMPLIFIER

DPM-900B Series, CDV/CDA-900A, MCF-B etc.

STATIC STRAIN MEASURING INSTRUMENTS

UCAM-80A etc.

ACQUISITION INSTRUMENT for dynamic strain

PCD-400B Series, CTRS-100 Series, EDX-5000A etc.

INSTRUMENTATION CONDITIONERS

WGA-680A, WGA-910A etc.

Kyowa Electronic Instruments provides various conditioners. We recommend WGA-680A and WGA-710C for your case. See here for more information.

Examples of measuring instruments supporting remote sensing are CDV-900A, WP-120A, and WGA-710C. When connecting a transducer supporting remote sensing to a measuring instrument supporting remote sensing, use the calibration factor on the Teat Data Sheet of the transducer. When using a measuring instrument not supporting remote sensing, the output drops due to the cable resistance, thus you need to use the calibration factor to compensate for the cable resistance.

The logic of the I/O ports of this product are NPN circuits. See the instruction manual for more information.
Download the instruction manual here.

Principle

You must have a dedicated Strain Measuring Instrument.
Since the sensor side forms a Wheat Stone Bridge, in theory, it is possible to measure strain as long as you have a power supply to apply the excitation voltage and a voltmeter to measure the change in voltage (potential difference). When considering factors such as the power supply’s stability and S/N ratio and the potential difference of output voltage, it becomes unrealistic.
The dedicated strain measuring instrument is designed with these factors in consideration, allowing for strain based accuracy adjustment for convenience.

Slow phenomena changes are called Static Strain and fast changing phenomenon is called Dynamic Strain.

There are two types of instruments for measuring the dynamic phenomena which are AC Power Bridge and DC Power Bridge. Characteristics of each are as follows
AC type DC type a (DPM-711B)

 

AC type (DPM-911B)

DC type (CDV-900A)

Bridge power voltage

AC (AC 2Vrms, 0.5Vrms)

Direct current (10, 4, 2, 1VDC)

Frequency range

DC to 2.5kHz

DC to 500kHz

Sensitivity (at theinput of 10με)

0.2V(BV=AC2Vrms)

0.1V(BV=DC2V)

SN ratio

60dBp-p (about 1000 times)
(Low pass filter = FLAT)

About 40dBp-p(about 100 times)
(BV = 2VDC, low pass filter = FLAT)

Effects of external inductive noise

Not received much

Susceptible

Balance adjustment

R balance
C balance (Automatic tracking at CST)

R balance Only

Temperature effect on zero balance

±0.1×10-6strain /℃ or less

±1×10-6strain /℃ or less

Temperature effect on sensitivity

±0.05% /℃ or less

±0.01% /℃ or less

Primary use

General strain measurement using strain gages

Measurements using the transducer,
strain measurement of high frequency

DC system with a signal conditioner is commonly used to apply a constant voltage to the bridge and when you extend the cable connection, sensitivity is reduced by cable resistance.
The sensitivity of the constant current method, on the other hand, does not decrease because it uses constant current, and according to the Ohm’s law (V=I・R), if the bridge resistant is constant, then the excitation voltage applied to the transducer remains constant.

There are two methods used when measuring multiple channels, the Scanning Method and Simultaneous Sampling method.
This method has been adopted by static strain measuring instrument to measure the slow-changing strain.
Simultaneous sampling method is a method to simultaneously measure all the channels at the same time, eliminating the time lag between channels.
This method has been adopted for the acquisition instrument form measuring the dynamic strain time series changes in the strain of a moment.

The calibration constant is the coefficient used to convert the strain value acquired by a strain gage transducer into physical quantity.
In the measurement condition setting for both static and dynamic strain measurement instruments, you can input the calibration constant, allowing you to acquire data in physical quantity.

The gage factor of the measuring instrument is fixed at 2. You can calculate the true strain (ε) from the measured strain (ε0) with the equation below.
(ε)=2.00/Ks×(ε0), Ks: Gage factor of gage used.

How to use

As basic measures there is single point ground. Single point grounding means to ground one of the series of the shielded part (Shield cable and metal housing), shielding it from external noise. Grounding 2 points from the same shielded series creates a potential difference causing the current to flow resulting in noise.
Please see other solutions here.

The DAS-200A data analysis software offers a file coupling function.
Two or more files recorded in a synchronous operation (master and slaves) can be coupled into a single file.